Our Calibration Services

TEMPERATURE
(- 196 and -30 to 600 ºC)
(-200 to +2000 °C with calibrator)

Calibrated instruments

  • Liquid in glass thermometers
  • All types of analog thermometers
  • Industrial mercury or red liquid reading thermometers
  • Digital thermometers with RTD, thermistor or thermocouples probes
  • PRT high accuracy RTD; we provide a complete curve & table as per ITS90.
  • Temperature probes alone.
  • Temperature transmitters
  • Infrared thermometers
  • Calibration baths
  • Temperature controllers and thermostats

Our standards

  • Hart 9101, 0 °C dry bath
  • Hart 9123 dry bath, from ambient up to 600 °C
  • Hart 7103 liquid bath, from -30 to 120 °C
  • Julabo liquid bath FK31 from -30 to 200°C
  • High accuracy Hart 1502A digital thermometers (3) with precision PRT probe
  • Altek 434 simulator/calibrator
  • Multimeter HP3458A.
  • Calibrator Fluke 5500A.
  • Black body for infrared thermometers EDL DBB350
  • Two temperature controlled chambers for measuring instruments that measure temperature by an element other than an immersion probe, from -10 to 50 °C.

 Accuracy details of our calibration capacities.

STANDARD RANGE ACCURACY INCERTITUDE
 -196°C and from -30 to 420°C
 

0.011 to -197°C
0.015 to -30°C

0.017 to 160°C
0.026 to 232°C
0.055 to 419°C

0.01°C (Triple point of water cell)
 
0.0002°C
29.7646°C
( Fixed cell melting point of gallium)
 
0.001°C
50 to 600°C
±0.1°C from 50 to 300°C
±0.5°C from 300 to 600°C
 

Infrared
-20 to 350°C
from 350 to 450°C

±0.25
±1°C

 