Description
All models measure either positive, negative or differential pressures with ±0.10% of full scale accuracy. You can select from up to seven common English and metric pressure units so conversions are not necessary. A memory function allows storage of up to 40 readings for later recall and a backlight provides auxiliary lighting for hard-to-see locations. Also standard is a hold feature; plus both visual and audible overpressure alarms.
- Service: Air and compatible gases.
- Accuracy: ±0.10% of full scale from 60 to 78F° (15.6 to 25.6°C); ±1% of full scale from 32 to 60 and 78 to 104°F (0 to 15 .6 and 25.6 to 40°C).
- Pressure Hysteresis:±0.1% of FS.Pressure
- Limits: See chart on catalog page.
- Temperature Limits: 0 to 140°F (-17.8 to 60°C).
- Compensated Temperature Limits: 32 to 104°F (0 to 40°C).
- Storage Temperature Limits:-4 to 176°F (-20 to 80°C).
- Display: 0.42" (10.6 mm) 4 digit LCD.
- Resolution: See chart on catalog page.
- Power Requirements: 9 V Alkaline battery, installed non-functional, user replaceable.
- Weight: 10.2 oz (289 g).
- Two barbed connections for use with 1/8" (3.18 mm) or 3/16" (4.76 mm) I.D. tubing for 477B-1, 477B-2, 477B-3, 477B-4 and 477B-5 only. Two compression fittings for use with 1/8" (3.18 mm) I.D. x 1/4" (6.35 mm) O.D. tubing for 477B-6 and 477B-7 only.
- Compliance: CE.