Handheld Digital Manometer

477B

The Series 477B Handheld Digital Manometers are versatile, hand-held, battery operated manometers available in several basic ranges from 0-20 in w.c. up to 100 psi.

Modele

  • Handheld Digital Manometer - 477B-1

    0 to 20" w.c. (0 to 4.982 kPa)

  • Handheld Digital Manometer - 477B-2

    0 to 40" w.c. (0 to 9.96 kPa)

  • Handheld Digital Manometer - 477B-3

    0 to 200" w.c. (0 to 49.82 kPa)

  • Handheld Digital Manometer - 477B-4

    0 to 10 psi (0 to 68.95 kPa)

  • Handheld Digital Manometer - 477B-5

    0 to 30 psi (0 to 206.9 kPa)

  • Handheld Digital Manometer - 477B-6

    0 to 50 psi (0 to 344.7 kPa)

  • Handheld Digital Manometer - 477B-7

    0 to 100 psi (0 to 689.5 kPa)

Description

All models measure either positive, negative or differential pressures with ±0.10% of full scale accuracy. You can select from up to seven common English and metric pressure units so conversions are not necessary. A memory function allows storage of up to 40 readings for later recall and a backlight provides auxiliary lighting for hard-to-see locations. Also standard is a hold feature; plus both visual and audible overpressure alarms.

  • Service: Air and compatible gases.
  • Accuracy: ±0.10% of full scale from 60 to 78F° (15.6 to 25.6°C); ±1% of full scale from 32 to 60 and 78 to 104°F (0 to 15 .6 and 25.6 to 40°C).
  • Pressure Hysteresis:±0.1% of FS.Pressure
  • Limits: See chart on catalog page.
  • Temperature Limits: 0 to 140°F (-17.8 to 60°C).
  • Compensated Temperature Limits: 32 to 104°F (0 to 40°C).
  • Storage Temperature Limits:-4 to 176°F (-20 to 80°C).
  • Display: 0.42" (10.6 mm) 4 digit LCD.
  • Resolution: See chart on catalog page.
  • Power Requirements: 9 V Alkaline battery, installed non-functional, user replaceable.
  • Weight: 10.2 oz (289 g).
  • Two barbed connections for use with 1/8" (3.18 mm) or 3/16" (4.76 mm) I.D. tubing for 477B-1, 477B-2, 477B-3, 477B-4 and 477B-5 only. Two compression fittings for use with 1/8" (3.18 mm) I.D. x 1/4" (6.35 mm) O.D. tubing for 477B-6 and 477B-7 only.
  • Compliance: CE.

 

Additional information

Modèle

477B-1, 477B-2, 477B-3, 477B-4, 477B-5, 477B-6, 477B-7

You may also like…