Product added to cart

An error has occurred. Please try again later or contact the site administrator.

Dew-Point Transmitter

Easidew EA2

The Easidew sensor is a moisture meter that measures both dew point and moisture content. It is simple to install as it is available with a wide range of process connections and electrical connectors. All Easidew dew-point transmitters are available with a service exchange program which reduces the cost and time spent on maintenance.s

Description

  • Measurement ranges -110 up to +20°Cdp
  • Dew point or ppmV moisture content output
  • Accuracy ±2 °Cdp
  • Traceable calibration certificate
  • Easidew I.S. IECEx, cQPSus, ATEX and UKCA approvals
  • Low cost of ownership and easy maintenance with sensor exchange program
  • 316 Stainless steel sensor and sample blocks
  • 5/8” UNF, 3/4” UNF, G 1/2” process connections
  • MiniDIN 43650, M12 electrical connectors
  • Oxygen Service cleaned
Easidew, Easidew M12 and Easidew 34 Transmitter Easidew I.S. Transmitter
Performance Specifications
Measurement range –110...+20°C (-166...+68°F) dew point; –100...+20°C (-148...+68°F) dew point; non standard ranges available on request
Accuracy ±2°C (±3.6°F) dew point*
Response time 5 mins to T95 (dry to wet)
Repeatability 0.5°C (0.9°F) dew point
Calibration Traceable 13-point calibration certificate
Electrical Specifications
Output signal 4...20 mA (2-wire connection, current source); User configurable over range
Easidew M12: Modbus RTU over RS485
Output Dew point or moisture content
Analog output scaled range Dew point: –110 up to +20°C (-166 up to +68°F); Moisture content in gas: 0 up to 3000 ppmV;
Supply voltage 12...28 V DC
Load resistance Max 250 Ω @ 14 V (500 Ω @ 24 V)
Current consumption 23 mA max, depending on output signal
Compliances CE & UKCA
UL61010-1 & CAN/CSA C22.2 No. 61010-1
Operating Specifications
Operating temperature -40...+60°C (-40...+140°F)
Compensated temperature range -20...+50°C (-4...+122°F)
Storage temperature -40...+60°C (-40...+140°F)
Operating pressure 45 MPa (450 barg/6527 psig) maximum
Flow rate 1...5 Nl/min mounted in standard sampling block; 0...10 m/sec direct insertion
Mechanical Specifications
Ingress protection IP66 in accordance with standard BS EN 60529:1992; NEMA 4 protection in accordance with standard NEMA 250–2003
Easidew M12: IP65
Intrinsically safe area certificates ATEX: II 1 G Ex ia IIC T4 Ga (-20...+70 °C / -4...+158°F)
IECEx: EX ia IIC T4 Ga (-20...+70 °C / -4...+158°F)
TR CU @ 0Ex ia IICT4 Ga (-20...+70 °C / -4...+158°F)
cQPSus: Class I, Division 1, Groups A, B, C & D, T4
Class I, Zone 0, AEx ia IIC T4 Gb, Ex ia IIC T4 Gb
Tamb +70 °C
Oxygen service Optional: cleaned for enriched oxygen
Housing material 316 stainless steel
Dimensions MiniDIN 43650 form C L=132mm x ø45mm (5.2” x ø1.77”) (with connector cable)
Easidew M12: M12 5 pin L = 155mm x ø45mm (6.1” x ø1.77”) (with connector cable)
Filter (sensor protection) Standard: HDPE <10μm
Optional: 316 stainless steel sintered guard <80μm
Process connection Easidew: 5/8” - 18 UNF
Easidew 34: 3/4” - 16 UNF
Easidew M12: 5/8” -18 UNF, 3/4” - 16 UNF, G1/2” BSP
5/8” -18 UNF, 3/4” - 16 UNF, G1/2” BSP
KF40 Flange Accessories Flange for 5/8” 18 UNF & 3/4” 16 UNF process connections
Weight 150g (5.29oz)
Electrical connections Easidew: MiniDIN 43650 form C
Easidew M12: M12 5 Pin (A coded)
MiniDIN 43650 form C
Mating electrical connectors Mating connector supplied as standard
Easidew M12: optional 0.8, 2, 5 metre (2.62, 6.56, 16.4 foot) M12 A coded connector/cable available
Diagnostic conditions (factory programmed) Conditions
Sensor fault: 23 mA
Under-range dew point: 4 mA
Over-range dew point: 20 mA
Approved galvanic isolators KFD2-CR-EX1.20200
KFD2-CR-EX1.30200
KFD0-CS-EX1.50P
KFD0-CS-EX2.50P
KFD2-STC4-EX1.H
MTL5041
MTL5040
NOTES * Over Compensated Temperature Range

Your cart

Your cart is empty.